6-6 Calibration NetScan Users Manual
Calibration of Main Unit
Offset Calibration of Main Unit
Required Equipment
-
Shorting bar, or 3 linked banana plugs (see following figure)
-
Calibration Card (installed in bottom slot, slot #1, of the main unit)
&$87,21
Do not install or remove any cards from the main unit, or from its expansion chassis
unless the unit is powered “OFF” and has no power sources connected to it.
&$87,21
Use approved ESD precautions, including static-free work area and grounded wrist
strap, when handling circuit boards and electronic components. Failure to do so
could cause equipment damage due to electrostatic discharge.