Strona 1 - Digital Thermometer

OMEGAHH503Digital Thermometer

Strona 2 - Servicing Europe:

81011124. When the thermometer is turned on. The Relative set value and Hi/Lo Limitsset value that was in use when thermometer was last turned off set

Strona 3 - SAFETY INFORMATION

9OPERATOR MAINTENANCEWARNINGTo avoid possible electrical shock, disconnect the thermocoupleconnectors from the thermometer before removing the cover.B

Strona 4 - SPECIFICATIONS

1HH 502/HH 503 CALIBRATION PROCEDURENote: The following calibration procedure should perform only by qualified technicianswho have access to the item

Strona 5 - ENVIRONMENTAL

WARRANTY / DISCLAIMEROMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 monthsfrom date

Strona 6 - OPERATING INSTRUCTIONS

Where Do I Find Everything I Need forProcess Measurement and Control?OMEGA...Of Course!TEMPERATUREþ Thermocouple, RTD & ThermistorProbes, Connecto

Strona 7

OMEGAnetSM On-Line Servicehttp://www.omega.comInternet [email protected] North America:USA: ISO 9001 Certified Canada:One Omega Drive, Box

Strona 8 - REL Relative mode

1INTRODUCTIONThis instrument is a 4½ digit, compact-sized portable digital thermometerdesigned to use external K-type and J-type thermocouples as temp

Strona 9

2SPECIFICATIONSELECTRICALTemperature Scale: Celsius or Fahrenheit user-selectableMeasurement Range:J-TYPE -200°C to 1050°C, (-328°F to 1922°F)K-TYPE -

Strona 10

ENVIRONMENTALAmbient Operating Ranges: 0°C to 50°C (32°F to 122°F) <80% R.H.Storage Temperature: -20°C to 60°C (-4°F to 140°F) <70% R.H.GENERALD

Strona 11 - OPERATOR MAINTENANCE

4213OPERATING INSTRUCTIONS Power SwitchThe key turns the thermometer on or off. In the data SET mode, can notpower off, must leave data SET m

Strona 12

5Cold Junction Temperature Display mode (second display)Press C.J. TEMP key to display the thermocouple input connector coldjunction temperature on se

Strona 13 - RETURN REQUESTS / INQUIRIES

67The true average of all the reading taken over at least 22 hours preiod canbe displayed. If 22 hours is exceeded, new averages are no longercalculat

Strona 14 - OMEGA...Of Course!

789Sec. Minu. Selecting the Time scaleReading the third displayed (Time) in either second or minute. When thethermometer turned on, it is set to secon